The bottom line

Strongest for trends, children and newborns

Electrical Cardiometry performs best where it is used most: in neonates and children, and in tracking change, such as the stroke volume response to a leg raise, fluid or a drug. Its average bias against reference methods is small. Like every non-invasive monitor, its single absolute readings in adults do not match invasive thermodilution closely: pooled error is about 47% across all non-invasive devices (Joosten 2017) and 42–48% for Electrical Cardiometry (Sanders 2020). That is why this academy teaches trends and responses to treatment rather than single numbers.

Correlation is not agreement

Two methods can rise and fall together (high r) yet differ by a litre a minute. Look for bias, limits of agreement and percentage error, not just r.

The 30% rule

A percentage error under about 30% is the usual threshold for interchangeability (Critchley and Critchley, 1999). Many bioimpedance studies land near or above it.

Trending is a different question

A monitor can be imprecise for single values yet useful for the direction of change, for example the stroke volume response to a passive leg raise.

Reviews and studies

From meta-analyses to the bedside

✓ supports the method · ~ mixed or descriptive · ≠ single readings not interchangeable with the invasive reference (a limit shared by all non-invasive monitors).

Neonates

StudyPatientsCompared withFindingVerdict
Noori et al., 201220 term newborns, 115 pairsEchocardiographyBias −4 mL/min; precision similar to echo; unaffected by a ductus✓
Grollmuss and Gonzalez, 201428 low and very-low-birth-weight infants, 228 pairsEchocardiographyWithin the 30% criterion✓
Boet et al., 201679 stable preterm infantsEchocardiographyGood correlation, variability similar to echo; reference curves✓
Hsu et al., 201736 preterm infants with a significant ductus, 105 pairsEchocardiographyPercentage error 28.6%: wide but clinically acceptable✓
Song et al., 201440 preterm infants, 108 pairsEchocardiographyPromising for trends, but poor at low output and affected by ventilation mode~

Children

StudyPatientsCompared withFindingVerdict
Norozi et al., 200832 children with congenital heart diseaseDirect Fickr = 0.97; acceptable agreement✓
Grollmuss et al., 201224 newborns after arterial switchDoppler echoPercentage error 29%: interchangeable✓
Rauch et al., 201364 obese childrenDoppler echor = 0.91, bias 0.015 L/min✓
Blohm et al., 201472 children in intensive careEchocardiographyPercentage error 26% overall; preterm infants outside limits~
Narula et al., 2017Children with structural heart diseasePulmonary artery catheterSmall bias, good reliability✓
Coté et al., 2015402 children under anaesthesiaNo reference methodDetected some events before other monitors; some readings were artefacts~
Awadhare et al., 2022Children with shock in the emergency departmentFluid responseA 15% change in the contractility index predicted fluid responsiveness (AUC 0.85)✓

Adults

StudyPatientsCompared withFindingVerdict
Schmidt et al., 200537 coronary surgery patientsTransoesophageal Dopplerr² = 0.86, bias 0.18 L/min✓
Zoremba et al., 2007Intensive care patientsPulmonary artery thermodilution and PiCCOPercentage error 26.5%✓
Heringlake et al., 200729 cardiac surgery patientsPulmonary artery thermodilutionLack of agreement; did not track changes≠
Cox et al., 201750 cardiac surgery patientsContinuous pulmonary artery catheterPercentage error over 30% throughout; poor trending≠
Greiwe et al., 202541 patients after bypass surgeryPulmonary artery thermodilutionPercentage error 47%, concordance 48%≠
Song et al., 202572 intensive care patientsPiCCOWide limits of agreement, limited trending≠
Raue et al., 200930 patients with sepsis or SIRSPiCCOPercentage error 54%≠

Obstetrics

StudyPatientsCompared withFindingVerdict
Hammad et al., 201960 women with preeclampsiaLung ultrasoundThoracic fluid content detected pulmonary oedema (AUROC 0.94)✓
Liu et al., 201542 caesarean deliveries under spinalCourse over spinal anaesthesiaOutput usually rose during spinal hypotension, informing vasopressor choice~
Feng and Liu, 202020 caesarean deliveriesTransthoracic echoPercentage error 62%; moderate trending≠

The big picture: pooled reviews

ReviewScopeFindingVerdict
Mansfield et al., 2022Paediatric meta-analysis, 25 studiesNo significant difference in mean values; acceptable in children and adolescents, uncertain in neonates and congenital heart disease~
van Wyk et al., 2025European neonatal consensus (ESPR)Useful for tracking change in neonates; not accurate enough for absolute output~
O'Neill et al., 202033 neonatal studies, about 2,000 infantsNarrative review of bioimpedance in neonates~
Sanders et al., 202024 studies of Electrical Cardiometry: 620 adults, 603 childrenBias near zero; percentage error 48% in adults and 42% in children, so single readings do not replace the reference methods≠
Joosten et al., 201737 studies, 1,543 patients, all non-invasive cardiac output devicesPooled percentage error 47% for the whole category: none is interchangeable with thermodilution≠
Guidelines and key methods

The framework the academy teaches within

  • Consensus on circulatory shock and hemodynamic monitoring (ESICM).Cecconi et al., Intensive Care Medicine 2014: dynamic over static measures; less invasive devices only where validated in shock · doi
  • Surviving Sepsis Campaign: international guidelines 2021.Evans et al., Intensive Care Medicine 2021: dynamic measures such as passive leg raise to guide fluids · doi
  • Surviving Sepsis Campaign: paediatric guidelines 2020.Weiss et al., Pediatric Critical Care Medicine 2020 · doi
  • Prediction of fluid responsiveness by passive leg raising: a meta-analysis.Monnet, Marik and Teboul, Intensive Care Medicine 2016: a 10% rise in output predicts response (sensitivity 0.85, specificity 0.91) · doi
  • Passive leg raising: five rules, not a drop of fluid!Monnet and Teboul, Critical Care 2015 · doi
  • The meta-analysis of cardiac output: the 30% criterion.Critchley and Critchley, Journal of Clinical Monitoring and Computing 1999 · doi
Research by Dr. Nassef

Electrical Cardiometry in his own work

  • Hemodynamic-guided therapy is associated with better hypertension control compared with standard care.Vascular Health and Risk Management, 2026 · 156 patients · first author · doi
  • Noninvasive cardiac output monitoring in septic shock patients: a retrospective study on hemodynamic status and outcomes.Medicine, 2025 · 50 patients · senior author · observational, no reference method · doi
  • The role of Electrical Cardiometry in paediatric and neonatal anaesthesia and intensive care: a narrative review.Anesthesia and Critical Care, 2022 · senior author · doi
  • A pilot crossover trial of helmet ventilation and exercise performance in chronic obstructive pulmonary disease.Scientific Reports, 2022 · doi
  • A pilot study of short-term hemodynamic effects of negative pressure ventilation in chronic obstructive pulmonary disease, assessed using electrical cardiometry.Annals of Noninvasive Electrocardiology, 2021 · senior author · doi