Part 1 · The standard monitor

Five numbers that tell you the end of the story

Heart rate, blood pressure, oxygen saturation, breathing rate and temperature have been on every monitor for decades. They are essential, but in early circulatory failure the body defends blood pressure by tightening vessels and speeding the heart, so pressure stays normal while flow is already falling. Hypotension is a late sign: the international shock consensus says it should not be required to diagnose shock.

0%10%20%30%40%

Drag to bleed a healthy adult, step by step. Watch which monitor notices first.

Standard monitor

Heart rate75bpm
Systolic BP120mmHg
SpO₂98%
No alarm

Flow monitoring

Stroke volume0% change
Cardiac output0% change
Vascular resistance0% change
No alarm

Teaching approximation built from: stroke volume falling about 2% for each 1% of blood lost (Hinojosa-Laborde 2014, doi); systolic pressure and pulse falling only after about a 55% fall in stroke volume in human volunteers (Ryan 2008, doi); stroke volume changing in 74% of donors after a 450 mL donation, heart rate in 65%, systolic pressure in 52% (Stewart 2014, doi); ATLS haemorrhage classes; ESICM shock consensus, Cecconi 2014 (doi). Alarm thresholds here are illustrative. Real patients vary widely: real-world trauma data show heart rate and pressure move even less than the textbook suggests (Mutschler 2013, doi).

ParameterWhat it tells youWhen it changes as the patient deterioratesLimitations
ECG and heart rateRhythm, ischaemia, rateRises from about 15–30% blood loss, often bluntedAffected by beta-blockers, pain, age, pacing. Irreplaceable for rhythm.
Blood pressurePerfusion pressureLate: after about 30% loss, or a 50–55% fall in stroke volumeHeld up by vasoconstriction; a poor guide to flow
SpO₂Oxygen saturationLate in shock; the signal can be lost when perfusion is poorSays nothing about flow or oxygen delivery. Irreplaceable for oxygenation.
Breathing rate, temperatureCompensation, infectionRises in moderate to severe lossOften poorly recorded; non-specific
Stroke volumeBlood ejected per beatEarliest: falls about 2% per 1% of volume lostElectrical Cardiometry's single values are imprecise; the trend matters most
Cardiac outputTotal flowIntermediate: the rising heart rate protects it at firstSame caveat as stroke volume
Vascular resistanceVessel tighteningRises early, then collapses just before decompensationCalculated from pressure and output, so it carries both errors
Thoracic fluidFluid in the chestRises with congestion and oedemaAffected by effusions and body habitus

The physiology, stroke volume falling well before pressure, is strong. Evidence that a flow monitor changes outcomes by detecting deterioration earlier is still limited and mostly observational. Flow monitoring adds to the standard monitor; it never replaces the ECG or SpO₂.

Part 2 · Invasive methods

Measuring flow: catheter or skin sensors?

Thermodilution through a pulmonary artery catheter, or through a central line and an arterial line (transpulmonary), has long been the reference for cardiac output. It comes with a cost to the patient that non-invasive monitoring avoids.

Pulmonary artery catheter (thermodilution)Transpulmonary thermodilution and pulse contourArterial pulse contour (uncalibrated)Electrical Cardiometry
What goes into the patientCatheter through a central vein and the heart into the pulmonary arteryCentral venous line plus an arterial line, usually femoralArterial lineNothing: four skin sensors
RisksArrhythmia, infection, bleeding, pneumothorax, thrombosis, rare pulmonary artery ruptureLine infection, bleeding, thrombosis, limb ischaemiaArterial line risksSkin irritation from the sensors
ReadingsIntermittent boluses; continuous only with special cathetersCalibrated bolus, then continuous estimate that drifts with vascular toneContinuous; less reliable when vessel tone changes fastContinuous, beat to beat
Time to first valueInsertion procedure, imaging, calibrationTwo lines and calibrationOnce the arterial line is inSeconds after the sensors go on
Newborns and small childrenRarely feasiblePossible in larger children, with paediatric cathetersPossible, but less validatedDesigned for all ages, including preterm babies
Where it can be usedIntensive care, cardiac theatreIntensive care, theatreWherever an arterial line existsICU, theatre, emergency, ward, clinic, transport
Skill neededSpecialist insertion and interpretationCentral and arterial accessArterial accessSensor placement in minutes
Accuracy of a single valueReference method, about ±20% itselfClose to the pulmonary catheterVaries with vascular toneWider spread than thermodilution; strongest for trends

The disadvantages of invasive monitoring, from the trials

No survival benefit

Randomised trials of the pulmonary artery catheter found no reduction in deaths: in high-risk surgery (Sandham 2003, NEJM), in intensive care (PAC-Man, Harvey 2005, Lancet) and in heart failure (ESCAPE 2005, JAMA). A Cochrane review confirmed no difference in mortality or length of stay (Rajaram 2013).

Real complications

In PAC-Man about one in ten catheterised patients had a complication, such as a haematoma at the insertion site or an arrhythmia. In ESCAPE, adverse events were roughly twice as common with the catheter. An earlier observational study linked right-heart catheterisation with higher mortality (Connors 1996, JAMA).

The reference has its own error

Thermodilution's own precision is about ±20%, which is why studies accept up to 30% disagreement between methods (Critchley 1999). It is also less reliable with tricuspid regurgitation (Hamilton 1989), intracardiac shunts and arrhythmias, and depends on bolus technique.

Too late, too narrow

A catheter is placed once a patient is already critically ill, in intensive care or theatre. It cannot screen the emergency department, follow a patient on the ward, or monitor a 900-gram baby. Non-invasive monitoring can start before the crisis.

So, is Electrical Cardiometry as accurate as thermodilution?

For a single absolute number, no: thermodilution remains the reference, and Electrical Cardiometry's individual readings spread more widely around it. Pooled studies put the difference at a 42–48% percentage error, similar to other non-invasive monitors (see the evidence). But part of that spread belongs to thermodilution itself, and bedside decisions mostly rest on change: did stroke volume rise after the leg raise, the fluid, the drug? That is where a continuous, risk-free monitor earns its place, and where no catheter can follow most patients.

The fair summary: use thermodilution when you need a reference value and the patient already justifies the lines; use Electrical Cardiometry to see trends early, in any patient, at any age, without a needle.